Abstract
Background
Volumetric-modulated Dynamic WaveArc therapy (VMDWAT) is a non-coplanar continuous volumetric modulated radiation therapy (VMAT) delivery technique. Here, we monitored mechanical errors and their impact on dose distributions in VMDWAT using logfiles throughout the course of treatment.
Methods
Fifteen patients were enrolled (2 skull base tumor patients and 13 prostate cancer patients). VMDWAT plans were created for the enrolled patients. The prescribed dose for the skull base tumor was set as 54 Gy at 1.8 Gy per fraction, and that for the prostate cancer was set as 72 to 78 Gy at 2 Gy per fraction. We acquired logfiles to monitor mechanical errors and their impact on dose distribution in each fraction. The root mean square error (RMSE) in the multi-leaf collimator (MLC), gantry angle, O-ring angle and monitor unit (MU) were calculated using logfiles throughout the course of VMDWAT for each patient. The dosimetric impact of mechanical errors throughout the course of VMDWAT was verified using a logfile-based dose reconstruction method. Dosimetric errors between the reconstructed plans and the original plans were assessed.
Results
A total of 517 datasets, including 55 datasets for the 2 skull base tumor patients and 462 datasets for the 13 prostate cancer patients, were acquired. The RMSE values were less than 0.1 mm, 0.2°, 0.1°, and 0.4 MU for MLC position, gantry angle, O-ring angle, and MU, respectively. For the skull base tumors, the absolute mean dosimetric errors and two standard deviations throughout the course of treatment were less than 1.4% and 1.1%, respectively. For prostate cancer, these absolute values were less than 0.3% and 0.5%, respectively. The largest dosimetric error of 2.5% was observed in a skull base tumor patient. The resultant dosimetric error in the accumulated daily delivered dose distribution, in the patient with the largest error, was up to 1.6% for all dose-volumetric parameters relative to the planned dose distribution.
Conclusions
MLC position, gantry rotation, O-ring rotation and MU were highly accurate and stable throughout the course of treatment. The daily dosimetric errors due to mechanical errors were small. VMDWAT provided high delivery accuracy and stability throughout the course of treatment.
Trial registration
UMIN000023870. Registered: 1 October 2016.
http://ift.tt/2EQt6PC
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου