Τετάρτη 26 Απριλίου 2017

Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system

Publication date: May–June 2017
Source:Reports of Practical Oncology & Radiotherapy, Volume 22, Issue 3
Author(s): Hideharu Miura, Shuichi Ozawa, Masahiro Hayata, Shintarou Tsuda, Tsubasa Enosaki, Kiyoshi Yamada, Yasushi Nagata
We report the characteristics of quality assurance (QA) image for Vero4DRT system with a kilo-voltage (kV) cone-beam computed tomography (CBCT) capability to perform image-guided radiation therapy (IGRT). To acquire a set of CBCT, the kV source is rotated either 215° clockwise (CW) (tube 1 from 5° to 220° and tube 2 from 275° to 130°) or counterclockwise (CCW) (tube 1 from 85° to 230° and tube 2 from 355° to 140°). Image geometry, image uniformity, high/low contrast resolutions, and contrast linearity were measured with a Catphan 504 CT phantom (The Phantom Laboratory, NY). The comparison between measured and expected distances shows an excellent agreement. The CBCT for Vero4DRT system cannot perform a full 360° rotation, which leads to a loss in uniformity for image acquisition. Separations were observed for high-contrast resolution, with eight line pairs per centimeter corresponding to a gap size of 0.063cm. For low-contrast resolution, the seventh largest hole was visible. This hole has a 4-mm diameter with 1.0% contrast level. We should check the contrast linearity compared with known value, even though it is out of range from the manufacturer manual.



http://ift.tt/2pmrBQc

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου