Assessment of the Monitor Unit Objective tool for VMAT in the Eclipse treatment planning system

elsevier-non-solus.png

Publication date: March–April 2018
Source:Reports of Practical Oncology & Radiotherapy, Volume 23, Issue 2
Author(s): Sara Jiménez-Puertas, David Sánchez-Artuñedo, Marcelino Hermida-López
AimThis work aims to achieve the highest possible monitor units (MU) reduction using the MU Objective tool included in the Eclipse treatment planning system, while preserving the plan quality.BackgroundThe treatment planning system Eclipse (Varian Medical Systems, Palo Alto, CA) includes a control mechanism for the number of monitor units of volumetric modulated arc therapy (VMAT) plans, named the MU Objective tool.Material and methodsForty prostate plans, 20 gynecological plans and 20 head and neck plans designed with VMAT were retrospectively studied. Each plan (base plan) was optimized without using the MU Objective tool, and it was re-optimized with different values of the Maximum MU (MaxMU) parameter of the MU Objective tool. MU differences were analyzed with a paired samples t-test and changes in plan quality were assessed with a set of parameters for OARs and PTVs.ResultsThe average relative MU difference (ΔMU¯) considering all treatment sites, was the highest when MaxMU=400 (−4.2%, p<0.001). For prostate plans, the lowest ΔMU¯ was obtained (−3.7%, p<0.001). For head and neck plans ΔMU¯ was −7.3% (p<0.001) and for gynecological plans ΔMU¯ was 7.0% (p=0.002). Although similar MU reductions were observed for both sites, for some gynecological plans maximum differences were greater than 10%. All the assessed parameters for PTVs and OARs sparing showed average differences below 2%.ConclusionFor the three studied clinical sites, establishing MaxMU=400 led to the optimum MU reduction, maintaining the original dose distribution and dosimetric parameters practically unaltered.



http://ift.tt/2HGxHks

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου