BACKGROUND: Radial arterial cannulation is most commonly done using palpation, but the use of ultrasound has increased the cannulation success rate. This improvement, albeit significant, has not led to a very high success rate especially in trainees. A modified ultrasound technique for vascular cannulation (dynamic needle tip positioning) has been described for peripheral venous cannulation. We therefore assessed the success rate of this technique compared to the palpation technique for radial artery cannulation in adult surgical patients. METHODS: We enrolled patients who were having nonemergent operations that required a radial arterial catheter for intraoperative monitoring. Patients were randomized to either palpation or dynamic needle tip positioning technique. Arterial cannulation was performed by anesthesia residents or faculty members. The primary end point was successful cannulation on the first pass. Secondary end points were overall 5-minute success rate and number of attempts within 5 minutes. RESULTS: Two hundred sixty patients were evaluated. The first-pass success rate was 83% in the dynamic needle tip positioning technique group (n = 132) and 48% in the palpation group (n = 128; P