Abstract
A new Optically Stimulated Luminescence Badge Reader (OSBARE-1) system has been designed and developed for application in the individual monitoring dosimetry. This badge reader system utilizes the 470-nm light of a blue LED for CW-OSL readout with the help of PMT photon counting module. The developed reader system can process four element 24 OSLD cards within 25 min. These four-element OSLD card consists of the Teflon embedded indigenously developed dosimetric grade α-Al
2O
3:C phosphor. The minimum measurable dose (MMD) was found to be ~26 μGy for these OSLD cards with reproducibility of ~1.12%. The various operational parameters such as variation in the dark counts, OSL scattering background counts and reproducibility have been studied in detailed for this reader system. The dosimetric studies performed on this developed reader system found to have a great potential for the OSLD-based large-scale personnel monitoring progr am for the radiation workers.