Κυριακή 19 Σεπτεμβρίου 2021

Fully Automated Measurement of Cochlear Duct Length From Clinical Temporal Bone Computed Tomography

xlomafota13 shared this article with you from Inoreader

Objectives/Hypothesis

To present and validate a novel fully automated method to measure cochlear dimensions, including cochlear duct length (CDL).

Study Design

Cross-sectional study.

Methods

The computational method combined 1) a deep learning (DL) algorithm to segment the cochlea and otic capsule and 2) geometric analysis to measure anti-modiolar distances from the round window to the apex. The algorithm was trained using 165 manually segmented clinical computed tomography (CT). A Testing group of 159 CTs were then measured for cochlear diameter and width (A- and B-values) and CDL using the automated system and compared against manual measurements. The results were also compared with existing approaches and historical data. In addition, pre- and post-implantation scans from 27 cochlear implant recipients were studied to compare predicted versus actual array insertion depth.

Results

Measurements were successfully obtained in 98.1% of scans. The mean CDL to 900° was 35.52 mm (SD, 2.06; range, [30.91–40.50]), the mean A-value was 8.88 mm (0.47; [7.67–10.49]), and mean B-value was 6.38 mm (0.42; [5.16–7.38]). The R 2 fit of the automated to manual measurements was 0.87 for A-value, 0.70 for B-value, and 0.71 for CDL. For anti-modiolar arrays, the distance between the imaged and predicted array tip location was 0.57 mm (1.25; [0.13–5.28]).

Conclusion

Our method provides a fully automated means of cochlear analysis from clinical CTs. The distribution of CDL, dimensions, and cochlear quadrant lengths is similar to those from historical data. This approach requires no radiographic experience and is free from user-related variation.

Level of Evidence

3 Laryngoscope, 2021

View on the web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου