Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer

elsevier-non-solus.png

Publication date: 1 February 2018
Source:International Journal of Radiation Oncology*Biology*Physics, Volume 100, Issue 2
Author(s): Mary Adjeiwaah, Mikael Bylund, Josef A. Lundman, Camilla Thellenberg Karlsson, Joakim H. Jonsson, Tufve Nyholm
PurposeTo investigate the effect of magnetic resonance system- and patient-induced susceptibility distortions from a 3T scanner on dose distributions for prostate cancers.Methods and MaterialsCombined displacement fields from the residual system and patient-induced susceptibility distortions were used to distort 17 prostate patient CT images. VMAT dose plans were initially optimized on distorted CT images and the plan parameters transferred to the original patient CT images to calculate a new dose distribution.ResultsMaximum residual mean distortions of 3.19 mm at a radial distance of 25 cm and maximum mean patient-induced susceptibility shifts of 5.8 mm were found using the lowest bandwidth of 122 Hz per pixel. There was a dose difference of <0.5% between distorted and undistorted treatment plans. The 90% confidence intervals of the mean difference between the dCT and CT treatment plans were all within an equivalence interval of (−0.5, 0.5) for all investigated plan quality measures.ConclusionsPatient-induced susceptibility distortions at high field strengths in closed bore magnetic resonance scanners are larger than residual system distortions after using vendor-supplied 3-dimensional correction for the delineated regions studied. However, errors in dose due to disturbed patient outline and shifts caused by patient-induced susceptibility effects are below 0.5%.



from Cancer via ola Kala on Inoreader http://ift.tt/2mfpeME
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου